User Contributed MET/CAL Procedure ============================================================================= INSTRUMENT: HP 5371A TIME INTERVAL ANALYZER VERIFICATION (REV 1.1) DATE: 02-May-94 AUTHOR: User Contributed REVISION: 1.1 ADJUSTMENT THRESHOLD: 80% NUMBER OF TESTS: 23 NUMBER OF LINES: 324 CONFIGURATION: Fluke 6060B ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F L W 1.002 HEAD Preliminary instructions 1.003 DISP Connect the HP 5371A to Port 1 of the Controller. 1.003 DISP 1.003 DISP Connect the HP 3325A EXT REF IN to the HP 5371A 1.003 DISP FREQUENCY STANDARD OUTPUT. Leave this connected 1.003 DISP for all tests. 1.003 DISP 1.003 DISP Connect the HP 3325A SIGNAL OUTPUT to the 1.003 DISP HP 5371A CHANNEL A input. 1.004 HEAD Setting the HP 5371A memory registers, Please Wait. 1.005 IEEE [TERM CR] 1.006 IEEE *RST 1.007 IEEE [D2000] #turn off the write protect. 1.008 IEEE IST;REG,1;PROT,OFF 1.008 IEEE IST;REG,2;PROT,OFF 1.008 IEEE IST;REG,3;PROT,OFF 1.008 IEEE IST;REG,4;PROT,OFF 1.008 IEEE IST;REG,5;PROT,OFF 1.008 IEEE IST;REG,6;PROT,OFF 1.008 IEEE IST;REG,7;PROT,OFF 1.009 IEEE MEAS;FUNC,FREQ;SOUR,A;MEAS;MSIZ,1;ARM,ISAM;SAMP;DEL,1 1.010 IEEE INP;SOUR,A;TRIG,MAN 1.011 IEEE INP;SOUR,B;TRIG,MAN 1.012 IEEE SMOD,SING 1.013 IEEE *SAV,1 1.014 IEEE MEAS;FUNC,FREQ;SOUR,B;*SAV,2 1.015 IEEE MEAS;SOUR,(A^B);INP;MOD,COM;*SAV,3 1.016 IEEE INP;SOUR,A;LEV,1;INP;SOUR,B;LEV,1;*SAV,4 1.017 IEEE INP;SOUR,A;LEV,-1;INP;SOUR,B;LEV,-1;*SAV,5 1.018 IEEE *RST 1.019 IEEE [D2000] 1.020 IEEE MEAS;FUNC,PEAK;BLOC,1;SMOD,SING 1.021 IEEE *SAV,6 1.022 IEEE MEAS;FUNC,PEAK 1.023 IEEE MEAS;SOUR,B 1.024 IEEE SAMP;CHAN,B 1.025 IEEE *SAV,7 1.026 HEAD {Frequency Range Test, Channel A} 1.027 IEEE [@3325]FU1 OF0VO AM45MV FR.125HZ 1.028 IEEE *RCL,1 1.029 IEEE [D1000] 1.030 IEEE [I] 1.031 IEEE [D3000] 1.032 MEME 1.033 MEME 0.125 1.034 MEMC HZ 0.004U 2.001 HEAD {Frequency Range Test, Channel B} 2.002 DISP Disconnect the HP 3325A from Channel A of the HP 5371A. 2.002 DISP 2.002 DISP Connect the HP 3325A SIGNAL OUTPUT to Channel B 2.002 DISP of the HP 5371A. 2.003 IEEE *RCL,2 2.004 IEEE [I] 2.005 IEEE [D3000] 2.006 MEME 2.007 MEME 0.125 2.008 MEMC HZ 0.004U 3.001 HEAD {Frequency Range Test, Channel B (Squarewave)} 3.002 IEEE [@3325]AM2VO FU2 3.003 IEEE [D2000] 3.004 IEEE REST 3.005 IEEE [I] 3.006 IEEE [D3000] 3.007 MEME 3.008 MEME 0.125 3.009 MEMC HZ 0.004U 4.001 HEAD {Frequency Range Test, Channel A (Squarewave)} 4.002 DISP Disconnect the HP 3325A from Channel B. 4.002 DISP 4.002 DISP Connect the HP 3325A SIGNAL OUTPUT to Channel A 4.002 DISP of the HP 5371A. 4.003 IEEE *RCL,1 4.004 IEEE REST 4.005 IEEE [I] 4.006 IEEE [D3000] 4.007 MEME 4.008 MEME 0.125 4.009 MEMC HZ 0.004U 5.001 HEAD {Frequency Range Test, Channel A and B (500 MHZ)} 5.002 DISP Disconnect the HP 3325A from the HP 5371A. 5.002 DISP 5.002 DISP Connect the Fluke 6062A RF OUTPUT to the HP 5371A 5.002 DISP Channel A input. 5.003 6060 500000000H 45mVP -D S 5.004 IEEE *RCL,3 5.005 IEEE [D1000] 5.006 IEEE REST 5.007 IEEE [I] 5.008 MEME 5.009 MEMC HZ 100U 6.001 HEAD {Frequency Range Test, 2Vp-p Level} 6.002 6060 500000000H 2VP -D S 6.003 IEEE [D1000] 6.004 IEEE REST 6.005 IEEE [I] 6.006 MEME 6.007 MEMC HZ 100U 7.001 HEAD {Signal Operating Range Test, 1 Volt Offset} 7.002 DISP Disconnect the Fluke 6062A from the HP 5371A. 7.002 DISP 7.002 DISP Connect the HP 3325A SIGNAL OUTPUT to Channel A of the 7.002 DISP HP 5371A 7.003 IEEE [@3325]FU1 FR10MH OF1VO 7.004 IEEE *RCL,4 7.005 IEEE [D1000] 7.006 IEEE REST 7.007 IEEE [I] 7.008 MEME 7.009 MEME 10000000 7.010 MEMC HZ 0.2U 8.001 HEAD {Signal Operating Range Test, -1 Volt Offset} 8.002 IEEE [@3325]OF-1VO 8.003 IEEE [D1000] 8.004 IEEE *RCL,5 8.005 IEEE REST 8.006 IEEE [I] 8.007 MEME 8.008 MEME 10000000 8.009 MEMC HZ 0.2U 9.001 HEAD {Channel A Auto Trigger Tests (200 MHZ)} 9.002 DISP Disconnect the HP 3325A from Channel A. 9.002 DISP 9.002 DISP Connect the Fluke 6062A RF OUTPUT to Channel A. 9.003 IEEE [@6060]FR200MZ AP70.7MV 9.004 IEEE *RCL,6 9.005 IEEE REST 9.006 IEEE [I] 9.007 MEME 9.008 MEME 0.1 9.009 MEMC V 0.04U 10.001 IEEE [@6060]AP707MV 10.002 IEEE REST 10.003 IEEE [I] 10.004 MEME 10.005 MEME 1 10.006 MEMC V 0.4U 11.001 HEAD {Channel B Auto Trigger Tests (200 MHZ)} 11.002 DISP Disconnect the Fluke 6062A from Channel A. 11.002 DISP 11.002 DISP Connect the Fluke 6062A to Channel B of the HP 5371A. 11.003 IEEE [@6060]AP70.0MV 11.004 IEEE *RCL,7 11.005 IEEE [I] 11.006 MEME 11.007 MEME 0.1 11.008 MEMC V 0.04U 12.001 IEEE [@6060]AP707MV 12.002 IEEE REST 12.003 IEEE [I] 12.004 MEME 12.005 MEME 1 12.006 MEMC V 0.4U 13.001 HEAD {Auto Trigger Tests, Channel A (1 KHZ)} 13.002 DISP Disconnect the Fluke 6062A from Channel B. 13.002 DISP 13.002 DISP Connect the HP 3325A SIGNAL OUTPUT to Channel A. 13.003 IEEE [@3325]FU1 13.004 IEEE [D1000] 13.005 IEEE [@3325]FR1KH AM200MV OF0VO 13.006 IEEE [D1000] 13.007 IEEE *RCL,6 13.008 IEEE [D1000] 13.009 IEEE REST 13.010 IEEE [I] 13.011 MEME 13.012 MEME 0.1 13.013 MEMC V 0.04U 14.001 IEEE [@3325]AM2VO 14.002 IEEE [D2000] 14.003 IEEE REST 14.004 IEEE [I] 14.005 MEME 14.006 MEME 1 14.007 MEMC V 0.4U 15.001 HEAD {Auto Trigger Tests, Channel B (1 KHZ)} 15.002 DISP Disconnect the HP 3325A from Channel A. 15.002 DISP 15.002 DISP Connect the HP 3325A SIGNAL OUTPUT to Channel B. 15.003 IEEE [@3325]AM200MV 15.004 IEEE *RCL,7 15.005 IEEE [D1000] 15.006 IEEE REST 15.007 IEEE [I] 15.008 MEME 15.009 MEME 0.1 15.010 MEMC V 0.04U 16.001 IEEE [@3325]AM2VO 16.002 IEEE [D2000] 16.003 IEEE REST 16.004 IEEE [I] 16.005 MEME 16.006 MEME 1 16.007 MEMC V 0.4U 17.001 HEAD Setting The HP 5371A Memory Registers, Please Wait. 17.002 IEEE [@3325]FR1HZ AM140MV OF0VO 17.003 IEEE *RST 17.004 IEEE MEAS;FUNC,FREQ;MSIZ,2;ARM,EDED;SAMP;CHAN,EXT 17.005 IEEE STAR;CHAN,EXT 17.006 IEEE NUM;EXP,ON;SMOD,SING 17.007 IEEE *SAV,1 17.008 IEEE INP;SOUR,EXT;LEV,2.5 17.009 IEEE *SAV,2 17.010 IEEE LEV,-2.5 17.011 IEEE *SAV,3 17.012 IEEE LEV,.1;MEAS;ARM,EDEV;SAMP;DEL,0 17.013 IEEE *SAV,4 17.014 IEEE MEAS;STAR;SLOPE,NEG 17.015 IEEE *SAV,5 17.016 HEAD {External Arm Test, 1 HZ} 17.017 DISP Disconnect the HP 3325A from Channel B. 17.017 DISP 17.017 DISP Connect the HP 3325A SIGNAL OUTPUT to the HP 5371A 17.017 DISP front panel EXTERNAL ARM input, using a 50 OHM 17.017 DISP FEEDTHRU TERMINATION. 17.018 DISP Connect a BNC TEE to the HP 5371A rear panel FREQUENCY 17.018 DISP STANDARD OUTPUT. 17.018 DISP 17.018 DISP Connect one side of the TEE to Channel A of the 17.018 DISP HP 5371A. 17.018 DISP The other side of the TEE should be connected to the 17.018 DISP HP 3325A rear panel EXT REF INPUT. 17.019 IEEE *RCL,1 17.020 IEEE [D1000] 17.021 IEEE REST 17.022 DISP The next sequence of tests will require you to 17.022 DISP type in the GATE time reading displayed on the 17.022 DISP HP 5371A CRT. 17.022 DISP 17.022 DISP You will be prompted when to enter the reading. 17.022 DISP 17.022 DISP NOTE: Use the upper GATE time reading displayed 17.022 DISP on the HP 5371A CRT. 17.023 DISP Press the HP 5371A LOCAL key, then the NUMERIC key. 17.023 DISP 17.023 DISP If the reading is in ms, ie 999.882 ms, enter it 17.023 DISP as .999882 17.024 IEEE [D2000] 17.025 MEMI Enter the GATE time reading displayed on the HP 5371A. 17.026 MEME 17.027 MEME 1 17.028 HEAD {External Arm Test, 1 HZ Squarewave} 17.029 MEMC sec 0.007U 18.001 IEEE [@3325]FU2 AM5VO 18.002 IEEE [D2000] 18.003 IEEE REST 18.004 IEEE [D2000] 18.005 MEMI Enter the GATE time reading displayed on the HP 5371A. 18.006 MEME 18.007 MEME 1 18.008 MEMC sec 0.007U 19.001 HEAD {External Arm Test, 100 MHZ} 19.002 DISP Disconnect the HP 3325A SIGNAL OUTPUT from the 19.002 DISP HP 5371A EXTERNAL ARM INPUT. 19.002 DISP 19.002 DISP Connect the Fluke 6062A RF OUTPUT to the HP 5371A 19.002 DISP EXTERNAL ARM INPUT, using a 50 OHM FEEDTHRU TERMINATION. 19.002 DISP 19.002 DISP When prompted, enter the reading exactly as it appears, 19.002 DISP i.e., type in 199.9 for a reading of 199.9 ns. 19.003 IEEE [@6060]FR100MZ AP-13DB 19.004 IEEE [D1000] 19.005 IEEE REST 19.006 IEEE [D2000] 19.007 MEMI Enter the GATE time reading displayed on the HP 5371A. 19.008 MEME 19.009 MEME 100 19.010 MEMC Sec 0.2U 20.001 HEAD {External Arm Input, Signal Operating Range Tests} 20.002 DISP Disconnect the Fluke 6062A from the EXTERNAL ARM INPUT. 20.002 DISP 20.002 DISP Connect the HP 3325A SIGNAL OUTPUT to the HP 5371A 20.002 DISP EXTERNAL ARM INPUT, using a 50 OHM FEEDTHRU TERMINATION. 20.002 DISP 20.002 DISP The timebases should still be locked together. 20.003 IEEE [@3325]FU1 FR10MH OF2.5VO AM5VO 20.004 IEEE [D2000] 20.005 IEEE *RCL,2 20.006 IEEE [D1000] 20.007 IEEE REST 20.008 IEEE [D2000] 20.009 MEMI Enter the GATE time reading displayed on the HP 5371A. 20.010 MEME 20.011 MEME 200 20.012 MEMC nSec 0.2U 21.001 IEEE [@3325]OF-2.5VO 21.002 IEEE *RCL,3 21.003 IEEE [D1000] 21.004 IEEE REST 21.005 IEEE [D2000] 21.006 MEMI Enter the GATE time reading displayed on the HP 5371A. 21.007 MEME 21.008 MEME 200 21.009 MEMC nSec 0.2U 22.001 HEAD {TIME BASE ACCURACY} 22.002 DISP Disconnect the cable from the HP 3325A rear panel 22.002 DISP EXT REF INPUT. 22.002 DISP 22.002 DISP Connect the HP 5371A FREQUENCY STANDARD OUTPUT 22.002 DISP to INPUT A of the HP 5335A counter. 22.003 IEEE [@5335] IN 22.004 IEEE [@5335]AZ1 FN1 RE 22.005 IEEE [D1000] 22.006 IEEE [@5335][I] 22.007 MEME 22.008 MEME 10000000 22.009 MEMC HZ 5U 23.001 HEAD [27][91]1mVERIFICATION IS COMPLETE![27][91]m 23.002 DISP 23.002 DISP Interval is 1 year. 23.003 END